IRAS Explanatory Supplement
IX. The Low-Resolution Spectra
C. Data Processing
Chapter Contents
| Authors
| References
Table of Contents |
Index | Previous Section
| Next Section
- The Database
- Processing the Individual Spectra
- Despiking
- Conversion to a Linear Scale
- Interpolation to a Standard Wavelength Grid
- Correction for Wavelength-Dependent Responsivity
- Correction for Cross-Scan-Dependent Responsivity
- Overall Responsivity
- Joining of the Two Spectrum Halves
- Averaging Spectra, Quality Checks
- Final Selection of Spectra
The spectrometer data consist of three types of data: (i)
uncorrected spectra with header information, (ii) calibration
tables, and (iii) administrative files. The spectra were extracted
out of the data stream whenever an hours-confirmed source with
a signal-to-noise ratio greater than 25 or a source specified
as "known source"
(Section V.D.4), crossed the spectrometer
aperture. Extractions were also made for designated calibration
sources. The spectra were linked to survey sources using index
association records produced by the hours-and weeks- confirmation
processors (Section V.D.7).
There were three types of calibration tables: (a) the responses
of the five detectors to the internal reference source flashes
and the intensities of the flashes as derived by the survey calibration
processor; (b) correction tables for the relative responsivity
as a function of position across the five detectors; (c) correction
tables for the relative responsivity as a function of wavelength.
The cross-scan and wavelength dependent responsivity tables were
derived from special observations (see Section IX.B).
Processing of the data always started with the raw data.
This allowed the correction procedures to be improved continuously
up to the time of production of the catalog. The major processing
steps are discussed below.
Single sample spikes with an amplitude greater than 8% caused
by multiplexer errors (Section IX.B) were removed
and replaced
by an interpolated value using a simple algorithm operating on
the raw data. Multiple-sample spikes were not removed but their
presence was noted so that the spectrum-half would be rejected
later (see Section IX.C.3).
A lookup table of 256 entries was used to convert the raw
data to voltages on a linear scale. A standard reconstruction
of the input-voltage to the high-pass filter from the measured
output voltage was carried out. An offset correction was reset
to zero whenever the sample voltage dropped below a specified
threshold because of the effects of the zero clamping
(Section IX.A.4).
For ease of processing, an interpolation was carried out
to a standard regular grid of angular positions in the dispersion
direction. Because the dispersion of the spectrograph changed
rapidly as a function of the angular position
(Fig. IX.A.1), the
wavelength values corresponding to the standard sample values
were not equidistant. Before the interpolation, allowance was
made for variations in the scan speed of the telescope. For the
spectra of the brighter sources (signal-to-noise ratio greater
than 10) the well-defined in-scan detector edges were in-scan
to center the spectrum. The centering correction reduced in-scan
errors to approximately 2", corresponding to approximately
0.03 µm in wavelength.
The interpolated samples were multiplied by a responsivity
table sampled at the same standard grid. There was a table for
each of the five detectors derived from observations of
-Tau
(Fig. IX.B.2).
Although the software allowed selecting a different
table for each of 16 regularly spaced cross-scan positions on
the detector, the evidence for a cross-scan variation of the
wavelength-dependent
gain was too weak to justify using this option.
Depending on the nominal cross-scan position of the source,
a correction (Fig. IX.B.1) was applied for
the decrease of responsivity
towards the edges of the detectors. This correction was the weakest
link in the calibration process because of the relatively large
uncertainty in the cross-scan position. The correction applied
is uncertain by up to 20%, although this uncertainty was decreased
by the process of joining the two spectrum-halves together (see
Section IX.C.2.g).
The overall responsivity depended on the individual detector,
on the time and/or on the sky position. To account for these
variations a correction was derived from the voltage responses
to the two internal reference source flashes bracketing the time
of observation. After applying the responsivity
correction to a large sample of spectra the integrated fluxes
in the spectra were compared to the fluxes measured by the survey
array. Systematic factors of 0.75 and 1.00 were applied to the
integrated spectrometer fluxes to bring them in line with the
survey observations.
The two spectrum-halves (8-12 µm and 11-25 µm) were treated
independently until this point. Because of uncertainties in the
cross-scan position of the scan path over the detectors and therefore
in the nominal cross-scan responsivity correction, the two spectrum-halves
often differed by up to 15 or 20% after subtraction of a linear
baseline. The overlapping portion of the spectrum-halves was
used to determine another correction factor. In doing so, the
nominal relative cross-scan positions were used to determine which
half of the spectrum to change by the largest amount. If either
half had been observed by the central part of a detector, it was
considered reliably calibrated, and that portion was not changed
by the joining process, and the half of the spectrum observed
near the edge of a detector was shifted up or down towards the
other half. If both halves were considered equally reliable, then
each was scaled by the square root of the ratio between the overlapping
sections. This joining process reduced the overall error in the
responsivity correction to less than 10%.
Before spectra were averaged, a number of quality checks
were performed on the individual measurements of the two halves
of a source's spectrum. First, all measurements made within 18"
of the edge of any detector were flagged. Measurements were rejected:
- if they contained multiple-sample spikes (Section IX.C.2);
- if the join-factor obtained before (Section IX.C.2)
was outside the range 0.30 to 3.3;
- if the measurement was confused by neighboring sources;
this was considered to occur when the measurement met one of two
criteria: (i) the central portion was below the baseline determined
from signal-free parts of the spectrum-half; (ii) the baseline
at the low wavelength end of the spectrum-half differed from that
at the high wavelength end by more than 20% of the signal in the
8-12 µm band or by 10% in the 11-25 µm band.
The lower limit of these thresholds was 2.5 times the sample noise.
- if the measurement did not correlate with the "reference
measurement", defined as the measurement with the smallest
number of check-flags. This choice gave preference to measurements
passing over the central part of the detector. Any 8-12 µm measurement
for which the correlation coefficient with the reference was below
60% or any 11-25 µm measurement for which it was below 50% was
rejected. For line spectra without a continuum in the 8-12 µm
region the first criterion was waived.
At least 80% of the spectra in the catalog had correlation
coefficients above 70 and 60% in the short and long wavelength
halves, respectively. Some 40% correlated internally with coefficients
better than 80% in both spectrum-halves (see Section IX.C.4).
The spectrum-halves passing through all of the above tests
were averaged using the inverse of the square of the noise as
a weighting factor. At least two measurements in each of the two
spectrum-halves (8-13 and 11-25 µm) had to be accepted
before the spectrum could
be averaged and included in the spectral catalog.
After averaging the two spectrum-halves were rejoined, giving
both halves equal weight (see Section IX.C.1). Generally the
join factors differed from 1.00 by only a few percent.
The averaged spectrum was convolved with the 12 µm survey
passband. The integrated flux thus obtained, was compared to the
average 12 µm survey flux of the source. The ratio between the
two fluxes is given in the low resolution spectrometer catalog
record and has a 1
dispersion around
unity of about 15%. Exceptions to this rule will be spectra with sharp
lines (classes 8 and 9; see Section IX.D.2) or
small 12 µm fluxes.
Three selection criteria were applied for inclusion in the
Catalog of Low Resolution Spectra.
- The source is contained in the IRAS point source catalog.
- The entire spectrum must have been observed at least
twice and the individual measurements should be mutually consistent.
(individual spectra must pass all the checks mentioned in
Section IX.C.3 and must have a minimum correlation
coefficient of 50% between any two measurements of the source. The
large majority of spectra had however, much higher correlation
coefficients (see below).
- The source must pass a subjective visual inspection.
About 2.5% of all sources were rejected by this process, mostly
because they showed non-point source characteristics or confusion
with other sources.
Four samples of sources were selected for inclusion in the
catalog.
- Sources whose 12 µm survey flux density was larger than
25 Jy or whose 25 µm survey flux density was larger than 50 Jy.
Individual spectrum-halves were required to correlate with each
other with a correlation coefficient of 80% in either spectrum-half.
This sample contains about 2150 sources.
- Sources with 12 µm flux densities larger than 1 Jy or 25
µm flux densities larger than 2 Jy but not contained in Sample
a. The vast majority of these sources proved to be brighter than
approximately 5 Jy at 12 µm or 10 Jy at 25 µm. The correlation
coefficients were required to exceed 70% in the 8-12 µm band and
60% in the 11-25 µm band. This sample contains about 2450 sources.
- Sources in the same flux density range as sample b but with
lower correlation coefficients: between 60 and 70% in the 11-22
µm spectrum-half or between 50 and 60% in the 11-25 µm hall;
respectively. There are roughly 850 sources in this sample.
- Sources with minimum survey flux densities of 1 Jy at 12
µm or 2 Jy at 25 µm. Of these sources only the 11-25 µm
spectrum-half was required to have been measured consistently with a minimum
correlation coefficient of 50% between individual measurements.
Out of this sample only sources with specified spectral lines
and not contained in sample a b, c were selected for the catalog.
The selection was carried out by the classification program: only
classes 8 and 9 (see Section IX.D.2 and
Table IX.D.1) were kept.
There are approximately 40 spectra with lines in this sample.
Samples a b, and c contain continuum sources and approximately
75% of the line sources of the catalog; sample d contains the
remainder of the line sources.
Chapter Contents
| Authors
| References
Table of Contents |
Index | Previous Section
| Next Section